Lab instrument installation and CLIA validation checklist

When a laboratory installs a new analyzer, CLIA requires the lab to confirm the FDA complexity of the exact test system, complete method verification or establishment under 42 CFR 493.1253 before reporting patient results, and put calibration, quality control, proficiency testing, procedures and qualified personnel in place. Form CMS-116 or a state license may also need a test-menu update.

Last reviewed Nov 24, 2026 · By the Sound Lab Consultants regulatory team · Sources listed below

Lab instrument installation and CLIA validation checklist
Lab instrument installation and CLIA validation checklist

A setup checklist that follows CLIA, not a sales brochure

Instrument manufacturers deliver the analyzer, reagents and instructions for use. The laboratory still has to complete method verification or establishment, write procedures, set QC and calibration verification, enroll in proficiency testing, confirm personnel qualifications and update Form CMS-116 or a state license when the test menu changes. This site walks through that work for each manufacturer group below. Record counts come from the FDA CLIA categorization download (clia_detail.zip, file dated November 23, 2026; 81,314 records with effective dates from 1993-07-26 through 2026-11-20; retrieved November 24, 2026, America/Phoenix).

Manufacturer groupFDA recordsUnique test systemsComplexity mix
Beckman Coulter8,73342923 waived, 8,150 moderate complexity, 560 high complexity
Roche Diagnostics7,697520219 waived, 7,120 moderate complexity, 358 high complexity
Siemens Healthineers6,475532275 waived, 5,578 moderate complexity, 622 high complexity
Abbott2,831346346 waived, 2,201 moderate complexity, 284 high complexity
Werfen (Instrumentation Laboratory)2,2902981,365 moderate complexity, 925 high complexity
QuidelOrtho2,234229113 waived, 1,976 moderate complexity, 145 high complexity
Thermo Fisher Scientific1,0381822 waived, 776 moderate complexity, 260 high complexity
Sysmex7829814 waived, 595 moderate complexity, 173 high complexity
Becton Dickinson (BD)60317126 waived, 199 moderate complexity, 378 high complexity
Nova Biomedical5538526 waived, 527 moderate complexity
Tosoh Bioscience49047485 moderate complexity, 5 high complexity
bioMérieux45012514 waived, 204 moderate complexity, 232 high complexity
Radiometer3867113 waived, 368 moderate complexity, 5 high complexity
Horiba32022320 moderate complexity
Bio-Rad Laboratories282712 waived, 113 moderate complexity, 167 high complexity
Mindray22312221 moderate complexity, 2 high complexity
Randox Laboratories20421186 moderate complexity, 18 high complexity
DiaSorin201372 waived, 163 moderate complexity, 36 high complexity
Revvity (PerkinElmer)18979189 high complexity
Luminex1883121 moderate complexity, 167 high complexity
Diatron17910178 moderate complexity, 1 high complexity
Cepheid1732614 waived, 139 moderate complexity, 20 high complexity
Diagnostica Stago1395759 moderate complexity, 80 high complexity
Sebia127369 moderate complexity, 118 high complexity
Polymedco1283014 waived, 111 moderate complexity, 3 high complexity
Sekisui Diagnostics1224722 waived, 99 moderate complexity, 1 high complexity
Hologic98485 moderate complexity, 93 high complexity
Erba Diagnostics74973 moderate complexity, 1 high complexity
Qiagen531035 moderate complexity, 18 high complexity
Abaxis (Piccolo)2416190 waived, 51 moderate complexity
Boule (Medonic)18318 moderate complexity
Cue Health111 waived

Every test system name on the manufacturer pages is copied from the FDA file. Sound Lab Consultants is independent of every manufacturer listed.

What CLIA requires when you install an analyzer

Complexity is assigned to the test system (the instrument plus the reagents and the intended use), not to the analyte alone. Confirm the categorization of the exact system in the FDA CLIA database. A test system that has not been categorized is treated as high complexity until it is categorized (42 CFR 493.17(c)(4)).

RequirementRuleWhat the lab must do before or as it starts patient testing
Method verification (unmodified FDA-cleared or approved system)42 CFR 493.1253(b)(1) Demonstrate accuracy, precision and reportable range comparable to the manufacturer's claims, and verify that the manufacturer's reference intervals are appropriate for the laboratory's patient population, before reporting patient results.
Establishment of performance specifications (modified systems or methods not FDA-cleared/approved)42 CFR 493.1253(b)(2) Establish accuracy, precision, analytical sensitivity, analytical specificity (interfering substances), reportable range, reference intervals and any other needed characteristic, then set calibration and control procedures from those data (42 CFR 493.1253(b)(3)).
Documentation42 CFR 493.1253(c) Record the actual measurements, reactions or observations and keep the verification or establishment report for the director's review and for surveyors.
Calibration and calibration verification42 CFR 493.1255 Calibrate following the manufacturer's instructions and materials (or equivalent criteria the lab has established). Perform calibration verification at least every 6 months, including a minimal, mid-point and maximum value, and after a complete change of reagents, major preventive maintenance, or when QC or other indicators so require, unless an exception in the rule applies.
Quality control (and IQCP when used)42 CFR 493.1256 Control procedures must monitor the complete analytic process. For most quantitative nonwaived tests that means two control materials at different concentrations each day of testing, unless a specialty rule or an approved equivalent (including a documented IQCP) applies. Establish acceptability criteria for every control lot (42 CFR 493.1256(d)(10)).
Procedure manual42 CFR 493.1251 A written procedure for each test, available to and followed by testing personnel, with the content 493.1251(b) requires, approved by the director.
Maintenance and function checks42 CFR 493.1254 Perform manufacturer-defined maintenance and function checks at least as often as the manufacturer specifies, and keep function checks within the manufacturer's limits before patient testing.
Proficiency testing42 CFR 493.801, 42 CFR 493.803 Enroll in an HHS-approved PT program for each specialty, subspecialty and regulated analyte the new system will report. Test PT samples like patient specimens. For tests not listed in Subpart I, verify accuracy at least twice a year (42 CFR 493.1236(c)).
Personnel by complexitySubpart M (42 CFR 493.1403–42 CFR 493.1495 as applicable) The director, technical consultant or technical supervisor, clinical consultant, general supervisor (high complexity) and testing personnel must meet the qualifications for the highest complexity on the new menu. Competency must be documented.
CMS-116 and certificate type42 CFR 493.51, 42 CFR 493.53, 42 CFR 493.39; Form CMS-116 Report changes in testing as the regulations and the state survey agency specify. Adding nonwaived testing to a Certificate of Waiver laboratory requires a new certificate type before those tests are reported. A new specialty or a higher complexity can also change fees and the survey path.
State licensureState law (varies) Some states require a state laboratory license, a test-menu amendment or a director permit in addition to CLIA. Check the state where the laboratory is located before going live.
Accreditation checklistsThe laboratory's AO standards Accredited laboratories must also meet the accrediting organization's method-verification, instrument and personnel checklist items, which must be at least as stringent as CLIA.

Do not report patient results on a newly installed nonwaived system until verification or establishment is complete and the director has reviewed the records.

Need a verification plan before go-live?

The full cluster on soundlabconsult.com starts at the lab instrument manufacturer CLIA consulting hub, with one page per manufacturer. Book a free consultation with Sound Lab Consultants, an independent consulting firm.

Manufacturer names and trademarks are the property of their respective owners. Sound Lab Consultants is an independent consulting firm and is not affiliated with, authorized by, certified by or endorsed by any manufacturer, the FDA, CMS, CAP, COLA, The Joint Commission or CDPH.

Talk to Sound Lab Consultants

Sound Lab Consultants provides clinical laboratory consulting services nationwide, from startup and licensing to quality systems, oversight and survey response.

Book a free consultation

Related guides on soundlabconsult.com:

Frequently asked questions

Who is responsible for CLIA validation, the manufacturer or the lab?

The laboratory. The manufacturer provides the cleared or approved test system and its instructions. 42 CFR 493.1253 requires the laboratory to verify or establish performance specifications before reporting patient results.

Where do the manufacturer lists come from?

The FDA CLIA categorization download (clia_detail.zip dated November 23, 2026, retrieved November 24, 2026). Pages quote unique TEST_SYSTEM_NAME values and complexity from that file.

Does Sound Lab Consultants sell these analyzers?

No. Sound Lab Consultants is an independent consulting firm and is not affiliated with any manufacturer.

Can you help with a manufacturer not listed here?

Yes. The CLIA rules are the same for any FDA-categorized test system. Book a free consultation.