Lab instrument installation and CLIA validation checklist
When a laboratory installs a new analyzer, CLIA requires the lab to confirm the FDA complexity of the exact test system, complete method verification or establishment under 42 CFR 493.1253 before reporting patient results, and put calibration, quality control, proficiency testing, procedures and qualified personnel in place. Form CMS-116 or a state license may also need a test-menu update.

A setup checklist that follows CLIA, not a sales brochure
Instrument manufacturers deliver the analyzer, reagents and instructions for use. The laboratory still has to complete method verification or establishment, write procedures, set QC and calibration verification, enroll in proficiency testing, confirm personnel qualifications and update Form CMS-116 or a state license when the test menu changes. This site walks through that work for each manufacturer group below. Record counts come from the FDA CLIA categorization download (clia_detail.zip, file dated November 23, 2026; 81,314 records with effective dates from 1993-07-26 through 2026-11-20; retrieved November 24, 2026, America/Phoenix).
| Manufacturer group | FDA records | Unique test systems | Complexity mix |
|---|---|---|---|
| Beckman Coulter | 8,733 | 429 | 23 waived, 8,150 moderate complexity, 560 high complexity |
| Roche Diagnostics | 7,697 | 520 | 219 waived, 7,120 moderate complexity, 358 high complexity |
| Siemens Healthineers | 6,475 | 532 | 275 waived, 5,578 moderate complexity, 622 high complexity |
| Abbott | 2,831 | 346 | 346 waived, 2,201 moderate complexity, 284 high complexity |
| Werfen (Instrumentation Laboratory) | 2,290 | 298 | 1,365 moderate complexity, 925 high complexity |
| QuidelOrtho | 2,234 | 229 | 113 waived, 1,976 moderate complexity, 145 high complexity |
| Thermo Fisher Scientific | 1,038 | 182 | 2 waived, 776 moderate complexity, 260 high complexity |
| Sysmex | 782 | 98 | 14 waived, 595 moderate complexity, 173 high complexity |
| Becton Dickinson (BD) | 603 | 171 | 26 waived, 199 moderate complexity, 378 high complexity |
| Nova Biomedical | 553 | 85 | 26 waived, 527 moderate complexity |
| Tosoh Bioscience | 490 | 47 | 485 moderate complexity, 5 high complexity |
| bioMérieux | 450 | 125 | 14 waived, 204 moderate complexity, 232 high complexity |
| Radiometer | 386 | 71 | 13 waived, 368 moderate complexity, 5 high complexity |
| Horiba | 320 | 22 | 320 moderate complexity |
| Bio-Rad Laboratories | 282 | 71 | 2 waived, 113 moderate complexity, 167 high complexity |
| Mindray | 223 | 12 | 221 moderate complexity, 2 high complexity |
| Randox Laboratories | 204 | 21 | 186 moderate complexity, 18 high complexity |
| DiaSorin | 201 | 37 | 2 waived, 163 moderate complexity, 36 high complexity |
| Revvity (PerkinElmer) | 189 | 79 | 189 high complexity |
| Luminex | 188 | 31 | 21 moderate complexity, 167 high complexity |
| Diatron | 179 | 10 | 178 moderate complexity, 1 high complexity |
| Cepheid | 173 | 26 | 14 waived, 139 moderate complexity, 20 high complexity |
| Diagnostica Stago | 139 | 57 | 59 moderate complexity, 80 high complexity |
| Sebia | 127 | 36 | 9 moderate complexity, 118 high complexity |
| Polymedco | 128 | 30 | 14 waived, 111 moderate complexity, 3 high complexity |
| Sekisui Diagnostics | 122 | 47 | 22 waived, 99 moderate complexity, 1 high complexity |
| Hologic | 98 | 48 | 5 moderate complexity, 93 high complexity |
| Erba Diagnostics | 74 | 9 | 73 moderate complexity, 1 high complexity |
| Qiagen | 53 | 10 | 35 moderate complexity, 18 high complexity |
| Abaxis (Piccolo) | 241 | 6 | 190 waived, 51 moderate complexity |
| Boule (Medonic) | 18 | 3 | 18 moderate complexity |
| Cue Health | 1 | 1 | 1 waived |
Every test system name on the manufacturer pages is copied from the FDA file. Sound Lab Consultants is independent of every manufacturer listed.
What CLIA requires when you install an analyzer
Complexity is assigned to the test system (the instrument plus the reagents and the intended use), not to the analyte alone. Confirm the categorization of the exact system in the FDA CLIA database. A test system that has not been categorized is treated as high complexity until it is categorized (42 CFR 493.17(c)(4)).
| Requirement | Rule | What the lab must do before or as it starts patient testing |
|---|---|---|
| Method verification (unmodified FDA-cleared or approved system) | 42 CFR 493.1253(b)(1) | Demonstrate accuracy, precision and reportable range comparable to the manufacturer's claims, and verify that the manufacturer's reference intervals are appropriate for the laboratory's patient population, before reporting patient results. |
| Establishment of performance specifications (modified systems or methods not FDA-cleared/approved) | 42 CFR 493.1253(b)(2) | Establish accuracy, precision, analytical sensitivity, analytical specificity (interfering substances), reportable range, reference intervals and any other needed characteristic, then set calibration and control procedures from those data (42 CFR 493.1253(b)(3)). |
| Documentation | 42 CFR 493.1253(c) | Record the actual measurements, reactions or observations and keep the verification or establishment report for the director's review and for surveyors. |
| Calibration and calibration verification | 42 CFR 493.1255 | Calibrate following the manufacturer's instructions and materials (or equivalent criteria the lab has established). Perform calibration verification at least every 6 months, including a minimal, mid-point and maximum value, and after a complete change of reagents, major preventive maintenance, or when QC or other indicators so require, unless an exception in the rule applies. |
| Quality control (and IQCP when used) | 42 CFR 493.1256 | Control procedures must monitor the complete analytic process. For most quantitative nonwaived tests that means two control materials at different concentrations each day of testing, unless a specialty rule or an approved equivalent (including a documented IQCP) applies. Establish acceptability criteria for every control lot (42 CFR 493.1256(d)(10)). |
| Procedure manual | 42 CFR 493.1251 | A written procedure for each test, available to and followed by testing personnel, with the content 493.1251(b) requires, approved by the director. |
| Maintenance and function checks | 42 CFR 493.1254 | Perform manufacturer-defined maintenance and function checks at least as often as the manufacturer specifies, and keep function checks within the manufacturer's limits before patient testing. |
| Proficiency testing | 42 CFR 493.801, 42 CFR 493.803 | Enroll in an HHS-approved PT program for each specialty, subspecialty and regulated analyte the new system will report. Test PT samples like patient specimens. For tests not listed in Subpart I, verify accuracy at least twice a year (42 CFR 493.1236(c)). |
| Personnel by complexity | Subpart M (42 CFR 493.1403–42 CFR 493.1495 as applicable) | The director, technical consultant or technical supervisor, clinical consultant, general supervisor (high complexity) and testing personnel must meet the qualifications for the highest complexity on the new menu. Competency must be documented. |
| CMS-116 and certificate type | 42 CFR 493.51, 42 CFR 493.53, 42 CFR 493.39; Form CMS-116 | Report changes in testing as the regulations and the state survey agency specify. Adding nonwaived testing to a Certificate of Waiver laboratory requires a new certificate type before those tests are reported. A new specialty or a higher complexity can also change fees and the survey path. |
| State licensure | State law (varies) | Some states require a state laboratory license, a test-menu amendment or a director permit in addition to CLIA. Check the state where the laboratory is located before going live. |
| Accreditation checklists | The laboratory's AO standards | Accredited laboratories must also meet the accrediting organization's method-verification, instrument and personnel checklist items, which must be at least as stringent as CLIA. |
Do not report patient results on a newly installed nonwaived system until verification or establishment is complete and the director has reviewed the records.
Need a verification plan before go-live?
The full cluster on soundlabconsult.com starts at the lab instrument manufacturer CLIA consulting hub, with one page per manufacturer. Book a free consultation with Sound Lab Consultants, an independent consulting firm.
Manufacturer names and trademarks are the property of their respective owners. Sound Lab Consultants is an independent consulting firm and is not affiliated with, authorized by, certified by or endorsed by any manufacturer, the FDA, CMS, CAP, COLA, The Joint Commission or CDPH.
Talk to Sound Lab Consultants
Sound Lab Consultants provides clinical laboratory consulting services nationwide, from startup and licensing to quality systems, oversight and survey response.
Related guides on soundlabconsult.com:
Frequently asked questions
Who is responsible for CLIA validation, the manufacturer or the lab?
The laboratory. The manufacturer provides the cleared or approved test system and its instructions. 42 CFR 493.1253 requires the laboratory to verify or establish performance specifications before reporting patient results.
Where do the manufacturer lists come from?
The FDA CLIA categorization download (clia_detail.zip dated November 23, 2026, retrieved November 24, 2026). Pages quote unique TEST_SYSTEM_NAME values and complexity from that file.
Does Sound Lab Consultants sell these analyzers?
No. Sound Lab Consultants is an independent consulting firm and is not affiliated with any manufacturer.
Can you help with a manufacturer not listed here?
Yes. The CLIA rules are the same for any FDA-categorized test system. Book a free consultation.
Sources
- FDA, Clinical Laboratory Improvement Amendments — Download Data (clia_detail.zip dated November 23, 2026; retrieved November 24, 2026)
- FDA CLIA test categorization database (search)
- 42 CFR 493.1253, Establishment and verification of performance specifications
- 42 CFR 493.1255, Calibration and calibration verification procedures
- 42 CFR 493.1256, Control procedures
- 42 CFR 493.801, Enrollment and testing of samples
- Form CMS-116, CLIA Application for Certification
- 42 CFR 493.51, Notification requirements