Laboratory Consulting ServicesClinical lab consulting, start to survey Book a free consultation
Home › Bio-Rad Laboratories analyzer installation and CLIA validation checklist

Bio-Rad Laboratories analyzer installation and CLIA validation checklist

A laboratory that installs a Bio-Rad analyzer must confirm the FDA complexity of the exact test system, complete method verification or establishment under 42 CFR 493.1253 before reporting patient results, and put calibration, quality control, proficiency testing, procedures and qualified personnel in place. Sound Lab Consultants is an independent firm that helps labs plan and document that work.

Last reviewed Nov 24, 2026 · By the Sound Lab Consultants regulatory team · Sources listed below

Bio-Rad Laboratories analyzer installation and CLIA validation checklist
Bio-Rad Laboratories analyzer installation and CLIA validation checklist

FDA-listed Bio-Rad systems

In the FDA CLIA categorization download (clia_detail.zip, file dated November 23, 2026; 81,314 records with effective dates from 1993-07-26 through 2026-11-20; retrieved November 24, 2026, America/Phoenix), this manufacturer group accounts for 282 categorization records across 71 unique test system names: 2 waived, 113 moderate complexity, 167 high complexity. Those are counts of FDA records, not sales or installed-base figures. Counts group the FDA-listed name variants shown on this page; they are not a claim of corporate identity for every historical name.

FDA test system nameComplexity in the fileAnalytesSpecialties
Bio-Rad IH-Reader 24High (28)10Immunohematology
Bio-Rad Medical Diagnostics GmbH IH-1000High (28)13Immunohematology
BIO-RAD LABORATORIES BIOPLEX 2200 MULTI-ANALYTE DETECTION SYSTEMModerate (25)13General Immunology, Virology
Bio-Rad IH-500High (25)11Immunohematology
Bio-Rad Laboratories BioPlex 2200 SystemModerate (24)15Endocrinology, General Immunology
Bio-Rad Laboratories, Inc. TOX/See Drug Screen TestModerate (12)12Toxicology/TDM
Bio-Rad RADIAS SystemModerate (12)12Endocrinology, General Chemistry, General Immunology
Bio-Rad Medical Diagnostics GmbH Tango infinityHigh (9)9Immunohematology
Bio-Rad QuantimuneHigh (6), Moderate (3)9Endocrinology, General Chemistry, Toxicology/TDM
Bio-Rad HPLCHigh (8)8Endocrinology, Toxicology/TDM
Bio-Rad CDM SystemHigh (7)7Endocrinology, General Chemistry, Toxicology/TDM
Bio-Rad CoTubeHigh (7)7Endocrinology
Bio-Rad Medical Diagnostics GmbH Tango Infinity TMModerate (6)6Immunohematology
Bio-Rad VARIANTModerate (3), High (1)4General Chemistry
BIO-RAD PLATELIA ASPERGILLUS EIAHigh (3)1Mycology

The table lists the 15 test system names with the most categorization records. The FDA file lists 71 unique test system names in this manufacturer group. Look up the exact analyzer and reagent in the FDA CLIA database before reporting patient results.

Installation and validation checklist for a Bio-Rad analyzer

  1. Confirm complexity. Look up the exact instrument and reagent in the FDA CLIA database. Uncategorized systems are treated as high complexity (42 CFR 493.17(c)(4)).
  2. Match the certificate. Waived-only certificates cannot add moderate or high complexity Bio-Rad assays until a new certificate type is issued. Personnel must meet Subpart M for the highest complexity on the new menu.
  3. Write a verification or establishment plan under 42 CFR 493.1253 and have the director approve it before patient testing.
  4. Collect data for accuracy, precision and reportable range (verification) or the full establishment set if the method is modified or not FDA-cleared.
  5. Set calibration, calibration verification and QC from 42 CFR 493.1255 and 42 CFR 493.1256, including lot acceptability criteria and, where used, an IQCP.
  6. Issue a procedure for each assay (42 CFR 493.1251) and train testing personnel; document competency.
  7. Enroll in PT for each new specialty, subspecialty and regulated analyte (42 CFR 493.801), or schedule twice-yearly accuracy verification for tests not in Subpart I (42 CFR 493.1236(c)).
  8. Schedule manufacturer maintenance and function checks (42 CFR 493.1254) and keep records.
  9. Update Form CMS-116 and any state license if the test menu, specialties or certificate type change (42 CFR 493.51, 42 CFR 493.53, 42 CFR 493.39).
  10. Director review of the verification report before the first patient result is released.

How Sound Lab Consultants helps

Sound Lab Consultants works with new and existing laboratories that are installing Bio-Rad analyzers, or adding assays to a system already on site. The firm is independent of the manufacturer. Typical work includes:

Work is available nationwide, remote or on site. No outcome is guaranteed; survey findings depend on the laboratory's own records and performance.

Planning a Bio-Rad install?

The full cluster on soundlabconsult.com starts at the lab instrument manufacturer CLIA consulting hub, with one page per manufacturer. Book a free consultation with Sound Lab Consultants, an independent consulting firm.

Manufacturer names and trademarks are the property of their respective owners. Sound Lab Consultants is an independent consulting firm and is not affiliated with, authorized by, certified by or endorsed by Bio-Rad Laboratories or any other manufacturer, the FDA, CMS, CAP, COLA, The Joint Commission or CDPH.

Talk to Sound Lab Consultants

Sound Lab Consultants provides clinical laboratory consulting services nationwide, from startup and licensing to quality systems, oversight and survey response.

Book a free consultation

Related guides on soundlabconsult.com:

Frequently asked questions

Is verification required for an FDA-cleared Bio-Rad analyzer?

Yes, for nonwaived testing. 42 CFR 493.1253(b)(1) requires verification of performance specifications before patient results are reported.

Where are Bio-Rad complexities listed?

In the FDA CLIA test categorization database. This page's table uses names and complexities from the November 23, 2026 FDA download.

Does Sound Lab Consultants work for the manufacturer?

No. Sound Lab Consultants is independent of Bio-Rad Laboratories and of the FDA, CMS, CAP, COLA, The Joint Commission and CDPH.

What if we already have the analyzer and are adding assays?

The same CLIA rules apply to each new test system or modification. Verification or establishment must be complete before those results are reported.

Sources

More on Laboratory Consulting Services