A laboratory that installs a Stago analyzer must confirm the FDA complexity of the exact test system, complete method verification or establishment under 42 CFR 493.1253 before reporting patient results, and put calibration, quality control, proficiency testing, procedures and qualified personnel in place. Sound Lab Consultants is an independent firm that helps labs plan and document that work.

Last reviewed Nov 24, 2026 · By the Sound Lab Consultants regulatory team · Sources listed below

Diagnostica Stago analyzer installation and CLIA validation checklist
Diagnostica Stago analyzer installation and CLIA validation checklist

FDA-listed Stago systems

In the FDA CLIA categorization download (clia_detail.zip, file dated November 23, 2026; 81,314 records with effective dates from 1993-07-26 through 2026-11-20; retrieved November 24, 2026, America/Phoenix), this manufacturer group accounts for 139 categorization records across 57 unique test system names: 59 moderate complexity, 80 high complexity. Those are counts of FDA records, not sales or installed-base figures. FDA records also list Tcoag test systems, a Stago company, when those names appear. Counts group the FDA-listed name variants shown on this page; they are not a claim of corporate identity for every historical name.

FDA test system nameComplexity in the fileAnalytesSpecialties
Diagnostica Stago STA R MaxModerate (10), High (19)21Hematology
STA Compact Max AnalyzerHigh (8), Moderate (5)13Hematology
Diagnostica Stago STA Satellite SystemHigh (5), Moderate (5)8Hematology
Diagnostica Stago ST4High (5), Moderate (4)9Hematology
Diagnostica Stago STA R Max 3High (2), Moderate (4)6Hematology
DIAGNOSTICA STAGO STA-CompactModerate (3), High (2)4Hematology
Diagnostica Stago STA Compact Max 3Moderate (4), High (1)5Hematology
Diagnostica Stago STA Satellite MaxModerate (4), High (1)5Hematology
DIAGNOSTICA STAGO STA-R EvolutionModerate (2), High (2)3Hematology
DIAGNOSTICA STAGO STACHROM PROTEIN CModerate (1), High (2)1Hematology
DIAGNOSTICA STAGO ASSERACHROM PF4High (2)1Hematology
DIAGNOSTICA STAGO STA-SatelliteModerate (2)2Hematology
Diagnostica Stago STA-R EvolutionModerate (1), High (1)2Hematology
DIAGNOSTICA STAGO ROTACHROM HEPARIN (LMW)Moderate (1)1Hematology
DIAGNOSTICA STAGO STA - LIA TEST FREE PROTEIN SModerate (1)1Hematology

The table lists the 15 test system names with the most categorization records. The FDA file lists 57 unique test system names in this manufacturer group. Look up the exact analyzer and reagent in the FDA CLIA database before reporting patient results.

Installation and validation checklist for a Stago analyzer

  1. Confirm complexity. Look up the exact instrument and reagent in the FDA CLIA database. Uncategorized systems are treated as high complexity (42 CFR 493.17(c)(4)).
  2. Match the certificate. Waived-only certificates cannot add moderate or high complexity Stago assays until a new certificate type is issued. Personnel must meet Subpart M for the highest complexity on the new menu.
  3. Write a verification or establishment plan under 42 CFR 493.1253 and have the director approve it before patient testing.
  4. Collect data for accuracy, precision and reportable range (verification) or the full establishment set if the method is modified or not FDA-cleared.
  5. Set calibration, calibration verification and QC from 42 CFR 493.1255 and 42 CFR 493.1256, including lot acceptability criteria and, where used, an IQCP.
  6. Issue a procedure for each assay (42 CFR 493.1251) and train testing personnel; document competency.
  7. Enroll in PT for each new specialty, subspecialty and regulated analyte (42 CFR 493.801), or schedule twice-yearly accuracy verification for tests not in Subpart I (42 CFR 493.1236(c)).
  8. Schedule manufacturer maintenance and function checks (42 CFR 493.1254) and keep records.
  9. Update Form CMS-116 and any state license if the test menu, specialties or certificate type change (42 CFR 493.51, 42 CFR 493.53, 42 CFR 493.39).
  10. Director review of the verification report before the first patient result is released.

How Sound Lab Consultants helps

Sound Lab Consultants works with new and existing laboratories that are installing Stago analyzers, or adding assays to a system already on site. The firm is independent of the manufacturer. Typical work includes:

Work is available nationwide, remote or on site. No outcome is guaranteed; survey findings depend on the laboratory's own records and performance.

Planning a Stago install?

The full cluster on soundlabconsult.com starts at the lab instrument manufacturer CLIA consulting hub, with one page per manufacturer. Book a free consultation with Sound Lab Consultants, an independent consulting firm.

Manufacturer names and trademarks are the property of their respective owners. Sound Lab Consultants is an independent consulting firm and is not affiliated with, authorized by, certified by or endorsed by Diagnostica Stago or any other manufacturer, the FDA, CMS, CAP, COLA, The Joint Commission or CDPH.

Talk to Sound Lab Consultants

Sound Lab Consultants provides clinical laboratory consulting services nationwide, from startup and licensing to quality systems, oversight and survey response.

Book a free consultation

Related guides on soundlabconsult.com:

Frequently asked questions

Is verification required for an FDA-cleared Stago analyzer?

Yes, for nonwaived testing. 42 CFR 493.1253(b)(1) requires verification of performance specifications before patient results are reported.

Where are Stago complexities listed?

In the FDA CLIA test categorization database. This page's table uses names and complexities from the November 23, 2026 FDA download.

Does Sound Lab Consultants work for the manufacturer?

No. Sound Lab Consultants is independent of Diagnostica Stago and of the FDA, CMS, CAP, COLA, The Joint Commission and CDPH.

What if we already have the analyzer and are adding assays?

The same CLIA rules apply to each new test system or modification. Verification or establishment must be complete before those results are reported.