Adding Total Protein to Your Lab: Setup and Validation Checklist
To add Total Protein testing, confirm the test system's complexity and your certificate, write the procedure, complete a verification study, set QC, calibration and PT or accuracy checks, define critical values, train staff, and get the director's sign-off before the first patient result.
Total Protein go-live checklist
- Pick the Total Protein method and test system for your volume, turnaround and specimen types.
- Look up its CLIA categorization in the FDA database and confirm your certificate covers it (42 CFR 493.17).
- Write the Total Protein procedure, including specimen requirements and rejection criteria.
- Run the verification study: accuracy, precision, reportable range, and reference interval check (42 CFR 493.1253(b)(1)). Establish full specifications if the test is modified or laboratory-developed ((b)(2)).
- Set the QC plan: at least two levels of QC each day of testing (unless an IQCP applies) (42 CFR 493.1256).
- Schedule calibration and six-month calibration verification (42 CFR 493.1255).
- Enroll in CMS-approved PT before or at go-live: Total Protein is a regulated analyte (routine chemistry, 42 CFR 493.931). CLIA PT acceptance limit: target value ±8%.
- Define Total Protein critical values with the director and the call procedure (42 CFR 493.1291(g)).
- Build and verify the LIS/interface result, units and reference intervals (42 CFR 493.1291(a)).
- Train staff and document initial competency before they report Total Protein results.
- Get the laboratory director's written approval before the first patient result.
Build these Total Protein pitfalls into your procedure
- Prolonged tourniquet time raises total protein.
- Plasma total protein includes fibrinogen and reads slightly higher than serum; keep specimen type consistent with reference intervals.
Related tests to plan with Total Protein
Full guide: Total Protein testing in the clinical lab: CLIA, PT and QC requirements on soundlabconsult.com, part of the Laboratory Test Compliance Library.
Talk to Sound Lab Consultants
Sound Lab Consultants provides clinical laboratory consulting services nationwide, from startup and licensing to quality systems, oversight and survey response.
Related guides on soundlabconsult.com:
Frequently asked questions
How do I validate a new Total Protein test?
For an unmodified FDA-cleared system, verify accuracy, precision and reportable range and check the reference interval before reporting patients (42 CFR 493.1253(b)(1)). Modified or laboratory-developed Total Protein tests need full validation under 493.1253(b)(2).
Do I need PT before offering Total Protein?
Total Protein is a regulated analyte (routine chemistry, 42 CFR 493.931). CLIA PT acceptance limit: target value ±8%.
Can Sound Lab Consultants set up Total Protein testing for my lab?
Yes. Sound Lab Consultants writes Total Protein verification plans, procedures and QC programs and helps labs go live. Book a free consultation.
Sources
- 42 CFR 493.931, Routine chemistry
- 42 CFR 493.17, Test categorization
- 42 CFR 493.1253, Establishment and verification of performance specifications
- 42 CFR 493.1255, Calibration and calibration verification procedures
- 42 CFR 493.1256, Control procedures
- 42 CFR 493.1291, Test report
- 42 CFR 493.1236, Evaluation of proficiency testing performance
- FDA CLIA test categorization database