Adding Rheumatoid Factor (RF) to Your Lab: Setup and Validation Checklist
To add Rheumatoid Factor testing, confirm the test system's complexity and your certificate, write the procedure, complete a verification study, set QC, calibration and PT or accuracy checks, define critical values, train staff, and get the director's sign-off before the first patient result.
Rheumatoid Factor go-live checklist
- Pick the Rheumatoid Factor method and test system for your volume, turnaround and specimen types.
- Look up its CLIA categorization in the FDA database and confirm your certificate covers it (42 CFR 493.17).
- Write the Rheumatoid Factor procedure, including specimen requirements and rejection criteria.
- Run the verification study: accuracy, precision, reportable range, and reference interval check (42 CFR 493.1253(b)(1)). Establish full specifications if the test is modified or laboratory-developed ((b)(2)).
- Set the QC plan: at least two levels of QC each day of testing (unless an IQCP applies) (42 CFR 493.1256).
- Schedule calibration and six-month calibration verification (42 CFR 493.1255).
- Enroll in CMS-approved PT before or at go-live: Rheumatoid Factor (RF) is a regulated analyte (general immunology, 42 CFR 493.927). CLIA PT acceptance limit: target value ±2 dilutions or positive or negative.
- Define Rheumatoid Factor critical values with the director and the call procedure (42 CFR 493.1291(g)).
- Build and verify the LIS/interface result, units and reference intervals (42 CFR 493.1291(a)).
- Train staff and document initial competency before they report Rheumatoid Factor results.
- Get the laboratory director's written approval before the first patient result.
Build these Rheumatoid Factor pitfalls into your procedure
- RF can interfere with other immunoassays; know where it affects your menu.
- Quantitative and titer methods report differently.
Related tests to plan with Rheumatoid Factor
Full guide: Rheumatoid Factor (RF) testing in the clinical lab: CLIA, PT and QC requirements on soundlabconsult.com, part of the Laboratory Test Compliance Library.
Talk to Sound Lab Consultants
Sound Lab Consultants provides clinical laboratory consulting services nationwide, from startup and licensing to quality systems, oversight and survey response.
Related guides on soundlabconsult.com:
Frequently asked questions
How do I validate a new Rheumatoid Factor test?
For an unmodified FDA-cleared system, verify accuracy, precision and reportable range and check the reference interval before reporting patients (42 CFR 493.1253(b)(1)). Modified or laboratory-developed Rheumatoid Factor tests need full validation under 493.1253(b)(2).
Do I need PT before offering Rheumatoid Factor?
Rheumatoid Factor (RF) is a regulated analyte (general immunology, 42 CFR 493.927). CLIA PT acceptance limit: target value ±2 dilutions or positive or negative.
Can Sound Lab Consultants set up Rheumatoid Factor testing for my lab?
Yes. Sound Lab Consultants writes Rheumatoid Factor verification plans, procedures and QC programs and helps labs go live. Book a free consultation.
Sources
- 42 CFR 493.927, General immunology
- 42 CFR 493.17, Test categorization
- 42 CFR 493.1253, Establishment and verification of performance specifications
- 42 CFR 493.1255, Calibration and calibration verification procedures
- 42 CFR 493.1256, Control procedures
- 42 CFR 493.1291, Test report
- 42 CFR 493.1236, Evaluation of proficiency testing performance
- FDA CLIA test categorization database