Adding Cell Identification (Blood Smear) to Your Lab: Setup and Validation Checklist
To add Cell Identification testing, confirm the test system's complexity and your certificate, write the procedure, complete a verification study, set QC, calibration and PT or accuracy checks, define critical values, train staff, and get the director's sign-off before the first patient result.
Cell Identification go-live checklist
- Pick the Cell Identification method and test system for your volume, turnaround and specimen types.
- Look up its CLIA categorization in the FDA database and confirm your certificate covers it (42 CFR 493.17).
- Write the Cell Identification procedure, including specimen requirements and rejection criteria.
- Run the verification study: accuracy, precision, reportable range, and reference interval check (42 CFR 493.1253(b)(1)). Establish full specifications if the test is modified or laboratory-developed ((b)(2)).
- Set the QC plan: at least two levels of QC each day of testing (unless an IQCP applies) (42 CFR 493.1256).
- Schedule calibration and six-month calibration verification (42 CFR 493.1255).
- Enroll in CMS-approved PT before or at go-live: Cell Identification (Blood Smear) is a regulated analyte (hematology, 42 CFR 493.941). CLIA PT acceptance limit: 80% or greater consensus on identification.
- Define Cell Identification critical values with the director and the call procedure (42 CFR 493.1291(g)).
- Build and verify the LIS/interface result, units and reference intervals (42 CFR 493.1291(a)).
- Train staff and document initial competency before they report Cell Identification results.
- Get the laboratory director's written approval before the first patient result.
Build these Cell Identification pitfalls into your procedure
- Smear quality (thickness, stain) drives identification errors; control stain QC.
- Morphology competency should be assessed for every person who reviews smears.
Related tests to plan with Cell Identification
- White Blood Cell Differential (full guide)
- Complete Blood Count (CBC)
- Leukocyte Count (WBC) (full guide)
Full guide: Cell Identification (Blood Smear) testing in the clinical lab: CLIA, PT and QC requirements on soundlabconsult.com, part of the Laboratory Test Compliance Library.
Talk to Sound Lab Consultants
Sound Lab Consultants provides clinical laboratory consulting services nationwide, from startup and licensing to quality systems, oversight and survey response.
Related guides on soundlabconsult.com:
Frequently asked questions
How do I validate a new Cell Identification test?
For an unmodified FDA-cleared system, verify accuracy, precision and reportable range and check the reference interval before reporting patients (42 CFR 493.1253(b)(1)). Modified or laboratory-developed Cell Identification tests need full validation under 493.1253(b)(2).
Do I need PT before offering Cell Identification?
Cell Identification (Blood Smear) is a regulated analyte (hematology, 42 CFR 493.941). CLIA PT acceptance limit: 80% or greater consensus on identification.
Can Sound Lab Consultants set up Cell Identification testing for my lab?
Yes. Sound Lab Consultants writes Cell Identification verification plans, procedures and QC programs and helps labs go live. Book a free consultation.
Sources
- 42 CFR 493.941, Hematology
- 42 CFR 493.17, Test categorization
- 42 CFR 493.1253, Establishment and verification of performance specifications
- 42 CFR 493.1255, Calibration and calibration verification procedures
- 42 CFR 493.1256, Control procedures
- 42 CFR 493.1291, Test report
- 42 CFR 493.1236, Evaluation of proficiency testing performance
- FDA CLIA test categorization database