Adding CA 125 (Cancer Antigen 125) to Your Lab: Setup and Validation Checklist

To add CA 125 testing, confirm the test system's complexity and your certificate, write the procedure, complete a verification study, set QC, calibration and PT or accuracy checks, define critical values, train staff, and get the director's sign-off before the first patient result.

Last reviewed Nov 24, 2026 · By the Sound Lab Consultants regulatory team · Sources listed below

CA 125 go-live checklist

  1. Pick the CA 125 method and test system for your volume, turnaround and specimen types.
  2. Look up its CLIA categorization in the FDA database and confirm your certificate covers it (42 CFR 493.17).
  3. Write the CA 125 procedure, including specimen requirements and rejection criteria.
  4. Run the verification study: accuracy, precision, reportable range, and reference interval check (42 CFR 493.1253(b)(1)). Establish full specifications if the test is modified or laboratory-developed ((b)(2)).
  5. Set the QC plan: at least two levels of QC each day of testing (unless an IQCP applies) (42 CFR 493.1256).
  6. Schedule calibration and six-month calibration verification (42 CFR 493.1255).
  7. Enroll in CMS-approved PT before or at go-live: CA 125 (Cancer Antigen 125) is a regulated analyte (endocrinology, 42 CFR 493.933). CLIA PT acceptance limit: target value ±20%.
  8. Define CA 125 critical values with the director and the call procedure (42 CFR 493.1291(g)).
  9. Build and verify the LIS/interface result, units and reference intervals (42 CFR 493.1291(a)).
  10. Train staff and document initial competency before they report CA 125 results.
  11. Get the laboratory director's written approval before the first patient result.

Build these CA 125 pitfalls into your procedure

Related tests to plan with CA 125

Full guide: CA 125 (Cancer Antigen 125) testing in the clinical lab: CLIA, PT and QC requirements on soundlabconsult.com, part of the Laboratory Test Compliance Library.

For laboratory compliance and operations only, not medical advice or patient care guidance.

Talk to Sound Lab Consultants

Sound Lab Consultants provides clinical laboratory consulting services nationwide, from startup and licensing to quality systems, oversight and survey response.

Book a free consultation

Related guides on soundlabconsult.com:

Frequently asked questions

How do I validate a new CA 125 test?

For an unmodified FDA-cleared system, verify accuracy, precision and reportable range and check the reference interval before reporting patients (42 CFR 493.1253(b)(1)). Modified or laboratory-developed CA 125 tests need full validation under 493.1253(b)(2).

Do I need PT before offering CA 125?

CA 125 (Cancer Antigen 125) is a regulated analyte (endocrinology, 42 CFR 493.933). CLIA PT acceptance limit: target value ±20%.

Can Sound Lab Consultants set up CA 125 testing for my lab?

Yes. Sound Lab Consultants writes CA 125 verification plans, procedures and QC programs and helps labs go live. Book a free consultation.