Adding Blood Gases (pH, pO2, pCO2) to Your Lab: Setup and Validation Checklist
To add Blood Gas testing, confirm the test system's complexity and your certificate, write the procedure, complete a verification study, set QC, calibration and PT or accuracy checks, define critical values, train staff, and get the director's sign-off before the first patient result.
Blood Gas go-live checklist
- Pick the Blood Gas method and test system for your volume, turnaround and specimen types.
- Look up its CLIA categorization in the FDA database and confirm your certificate covers it (42 CFR 493.17).
- Write the Blood Gas procedure, including specimen requirements and rejection criteria.
- Run the verification study: accuracy, precision, reportable range, and reference interval check (42 CFR 493.1253(b)(1)). Establish full specifications if the test is modified or laboratory-developed ((b)(2)).
- Set the QC plan: QC under 42 CFR 493.1267 (a control each 8 hours of testing, with low and high values each day) (42 CFR 493.1256).
- Schedule calibration and six-month calibration verification (42 CFR 493.1255).
- Enroll in CMS-approved PT before or at go-live: Blood Gases (pH, pO2, pCO2) is a regulated analyte (routine chemistry, 42 CFR 493.931). CLIA PT acceptance limit: pH: target value ±0.04; pCO2: target value ±8% or ±5 mm Hg (greater); pO2: target value ±15% or ±15 mmHg (greater).
- Define Blood Gas critical values with the director and the call procedure (42 CFR 493.1291(g)).
- Build and verify the LIS/interface result, units and reference intervals (42 CFR 493.1291(a)).
- Train staff and document initial competency before they report Blood Gas results.
- Get the laboratory director's written approval before the first patient result.
Build these Blood Gas pitfalls into your procedure
- Air bubbles left in the syringe shift pO2 and pCO2; expel air and cap immediately.
- Delayed analysis lets cells keep metabolizing; follow your analysis-time and storage rules.
- Incorrect heparin amount or liquid heparin dilution affects results; use the collection devices your procedure specifies.
Related tests to plan with Blood Gas
Full guide: Blood Gases (pH, pO2, pCO2) testing in the clinical lab: CLIA, PT and QC requirements on soundlabconsult.com, part of the Laboratory Test Compliance Library.
Talk to Sound Lab Consultants
Sound Lab Consultants provides clinical laboratory consulting services nationwide, from startup and licensing to quality systems, oversight and survey response.
Related guides on soundlabconsult.com:
Frequently asked questions
How do I validate a new Blood Gas test?
For an unmodified FDA-cleared system, verify accuracy, precision and reportable range and check the reference interval before reporting patients (42 CFR 493.1253(b)(1)). Modified or laboratory-developed Blood Gas tests need full validation under 493.1253(b)(2).
Do I need PT before offering Blood Gas?
Blood Gases (pH, pO2, pCO2) is a regulated analyte (routine chemistry, 42 CFR 493.931). CLIA PT acceptance limit: pH: target value ±0.04; pCO2: target value ±8% or ±5 mm Hg (greater); pO2: target value ±15% or ±15 mmHg (greater).
Can Sound Lab Consultants set up Blood Gas testing for my lab?
Yes. Sound Lab Consultants writes Blood Gas verification plans, procedures and QC programs and helps labs go live. Book a free consultation.
Sources
- 42 CFR 493.931, Routine chemistry
- 42 CFR 493.17, Test categorization
- 42 CFR 493.1253, Establishment and verification of performance specifications
- 42 CFR 493.1255, Calibration and calibration verification procedures
- 42 CFR 493.1256, Control procedures
- 42 CFR 493.1291, Test report
- 42 CFR 493.1236, Evaluation of proficiency testing performance
- FDA CLIA test categorization database